Tektronix 514: When are you from?

Here’s some of my notes about trying to understand a little bit more about the 514, in particular, mine (S/N 948)

The usual questions are:

  • When were they made and for how long?
  • how many different variants were there, and which were visibly different vs which ones were just internal revisions.
  • When is mine from, and which are the correct schematics?  Usually the schematics will indicate a range of serial numbers for which a particular schematic sheet is valid for.

Some of the things I observed in the manuals:

  • the schematics in the BAMA manual for this manual aren’t just raw scans, they’ve been re-composited.  It’s evident elsewhere in the manual as well.  Generally whoever put together the PDF manual did a nice job, but it appears they cropped out the serial number info when assembling the PDF schematic pages.  d’oh.
  • There’s a few pictures in the BAMA manual:
    1. TYPE 514 D CATHODE RAY OSCILLOSCOPE
                          SERIES “A”
      With the “serial” field on the bottom left of the unit.  Mine like this, except without the ‘SERIES “A”’ text. (no S/N shown next to “Serial” callout)
    2. TYPE 514AD CATHODE RAY OSCILLOSCOPE
      with the serial directly under that. (S/N #5181)
  • Comparing the BAMA schematics to the paper one I bought, I can confirm that the BAMA manual schematics are for S/N 3150 – 3408.
  • There’s a note in my manual that says:
    “Major circuit changes occurred in the Type 514AD oscilloscope at S/N 3409. Numbers in parenthesis indicate the earlier values in the Type 514D oscilloscope.”
    So the ‘A’ version is anything after 3409.  OK great.
  • But, looking at the versions of the vertical amplifier schematic, there is:
    1. “Type 514 / 514D Oscilloscope” S/N 101 – 3149.
    2. “Type 514 Series A Cathode Ray Oscilloscope”, S/N 3150-3408 (same one from BAMA) dated March ’53,
    3. “Type 514 D Cathode Ray Oscilloscope” S/N 3409+, dated March ’56, which has notes about changes in S/N 3650+
  • So it seems like there’s the 514, the 514 series A, and the 514A, all available with or without the ‘D’ for delay line.  The ‘Series A’ seems like it might be the precursor to the plain-old ‘A’ suffix, common of later models, but there’s some inconsistencies in naming conventions across the versions of schematics, so it’s difficult to be certain.  These are the earliest years of the company, so it’s not surprising to see evolution in the documentation.

Observations looking through the catalogs:

  • October ’50 514D is listed for $950.
  • August ’51 514D is listed for $950.  no ‘D’ in silkscreen.
  • March ’52: 514D is listed for $950.  Underside photo shows PS caps identical to mine, but ‘D’ looks like a part of the silkscreen.
  • March ‘53: catalog says there was a 514-D, but there’s no page for it.
  • August ’54: there was -AD only
  • August ’55: gone.

The punchline:

  • Made from 1950 – 1956
  • Mine is from somewhere between ’50 & ’52.  Had to guess I’d say ’51.
  • There were a few different milestones:
    1. S/N 101 – 3149
    2. S/N 3150 – 3408
    3. S/N 3409 – at least 5181
  • some minor changes within the first few hundred are called out in the schematics.
  • They weren’t advertised after ’53, but available until at least ’56.
  • BAMA schematics are incomplete (I’ll upload mine, promise)
  • I wonder if there were examples where the delay line was added post-sale, and the ‘D’ was hand-stamped in, like mine?

Tektronix 620 XY monitor repair

This is a wholly unremarkable early 80’s XY CRT monitor.  It’s like, the least interesting thing in my collection, but I’m taking a workshop in vector graphics next weekend, so I figured I’d pull this out of the ‘upstate home for wayward oscilloscopes’ and see if it worked.

IMG_5269.JPG

It did not.  I got a full intensity spot on the center of the screen, with no ability to move it around or adjust it’s brightness.  I bought the manual from Artek Manuals and off I went.

I actually spent way too much time on this repair; learning how the power supply worked, re-drawing schematics, and making detailed measurements.  I’ll spare you all of that.  It was a bunch of tantalum capacitors.  In gear of this era, it’s a very common mode of failure.  They usually fail shorted, so they’re easy to spot with a resistance measurement in situ.  A shorted capacitor can mean bad news, and tantalums have a reputation for failing violently, spewing their gooey capacitance all over the innards of your gear.  Fortunately, this is Tektronix, so the power supply has a few different layers of protection to minimize the collateral damage caused by single part failure.  The problem manifested as a low voltage coming from the 15v supply.  Under no load, this would float around 19v, but with a modest load, of 100Ω or so, it would happily regulate. The amplifier board had some shorted tantalum capacitors (C397, C398, C401, C402), which brought the load way down to a few ohms.  Fortunately, there’s a clever little current regulator in power supply that drops the voltage if there’s a near-short like this.

Replacing these got the 15v supply working, but uncovered/caused some issues in components elsewhere, notably caps in the -20V & -70V unregulated supplies.  I also inadvertently roached one of the capacitors while testing it, and a probe slip took out a diode.  It was… not my proudest moment.  There’s also a few fuses that saved my stupidity from making even more of a mess.

Here’s some shots of the re-worked area:

With C142 pulledIMG_5267.JPG

 

Replaced C142, CR142, CR143, C143.  IMG_5268.JPG

 

I brought it up on the variac and was rewarded with a controllable trace that responded well to inputs, waiting for some vector graphic goodness.  I smell a clock project in my future.

I will say, I’m used to working on the giant old tube gear, which is simply a joy to service.  This on the other hand, was not Tektronix’s brightest moment in industrial design.  Getting this thing apart felt needlessly belabored.  Just getting the boards disconnected from eachother was a giant hassle because of how one of the cables was run and how a shield just overlapped the board enough to get in it’s way.
IMG_5188.JPG

That’s the power supply on the top, the amplifier board on the bottom, and the connector in the center of the image.  There’s a line voltage cable pinched between the amp board, and a transformer shield, which I had to loosen to get the two apart. To get the amp board out, you had to remove the whole back, which of course the CRT was attached to.
IMG_5245.JPG

Also, no pilot light?  This was by far my least favorite piece of tek gear to work on thus far.  I’m just glad I got it working and off my bench.

Here’s the compulsory Lissajous pattern.  There’s some DC offset, but I think that’s the Waveforms audio oscillator.

IMG_5288.JPG

Tektronix 181 Time Mark Generator repair

I’d been in need of a Time Mark Generator for a while, and finally found one that was the right mix of price and form-factor.  With a few exceptions, I’m loathe to spend over $100 on a piece of old test gear, and the only units under that were either RM500 series units which I’ve been avoiding, or the 180 which is huge.  Speaking of the 180, Richard Sears has a nice write-up on both the 180 & 181.

181_PC_ISO

This is clearly an early unit, based on the 3 digit serial number, and the style of the enclosure.  Later models have the familiar textured aluminum side panels and rounded corners, while this bears a resemblance to the earliest of Textronix scopes like the 511, 514, etc.. I think this is also the first piece of tek gear I own thats the ‘wrong’ color.

After replacing the fuse holder, which required some gentle modification of the chassis (hooray drill press), I brought it up on the TU-75 variac, and got a trace.  After being on for 20 minutes or so, the slower markers became unstable and impossible to calibrate, and it would get worse with time.

I checked the power supply and found that the -150V supply was low, at around -139V (or, um, high, absolutely speaking).  The thing is, it would regulate this incorrect voltage like a champ.  Bringing it up on a variac, that -139V was stable from ~80V up to my uncomfortably hot line voltage of 126V.  But, over 20 minutes or so, that would slowly fall to -137V, which is where things would get unstable.

During our semi-annual encounter where he enables my oscilloscope addiction, Kurt suggested I suspect the caps around the feedback loop of the regulator, and he was dead right.  Those .01uF bumblebees were leaky over 50v.  They’re still good below that, so I could probably sell them to some audiot, insisting that the leakage is a part of their distinct warmth and tone.

Screen Shot 2015-11-14 at 7.42.43 PM

Lifted one side to testIMG_5171.JPG

Their tiny replacements
IMG_5174.JPG

And here’s the trace on my 453IMG_5105.JPG

This is the highest output, 10Mhz.  It’s a frequency multiplier, and I think that modulation is fixable in calibration.

IMG_5120.JPG

Here’s a pic of with the ‘check count’ switch enabled, showing all of the pulses.  If you notice, there’s 9 small pulses in between every major pulse, not counting the ‘half pulse’ right before the next major pulse.  That means the divide by 10 multi-vibrator is working properly.  This is on my Tek 551 before calibration, which is why it appears to be running fast.

IMG_5132.JPG

 

Here’s the 100µS output on my 7D20.  Spot on.

IMG_5154.JPG

A quick overview of how it works:

At the heart of the unit is crystal controlled oscillator.  Mine has the oven controlled option, which means there’s actually a little container with the crystal, a heater, and a bimetallic thermostat to keep the constant temperature, which minimizes frequency drift.

Here’s a few shots of the crystal oven:

IMG_5095.JPG

With the outer shell removedIMG_5097.JPG

With the top off, revealing the 1Mhz crystal.  IMG_5098.JPG

After the oscillator, and the 1Mhz pulse shaper, the signal is applied to a series of cascaded 10:1 frequency dividers, which give the 10µS – 10mS pulse outputs.  In more modern times, you would do this by counting the pulses, letting every 10th through.  These are actually re-triggerable multi-vibrators, with a hold-off time adjusted to only let every 10th pulse through, so it’s not actually counting, it’s waiting.  You can see this when you adjust the trim settings on the front panel.  If it’s incorrectly set, it will hold off for too few or too many pulses, so for example on the 10µS output, you’ll end up with 9µS or 11µS pulse spacing.  Using that ‘check count’ switch, this is easy to observe.

More info and pictures on the wiki

Tektronix 551: Adventures in timebase calibration

Relatively speaking, the 551 is one of the less useful scopes in my collection, as specially considering it’s size.  With a separate power supply at 42 lbs, and the mainframe 10 lbs over that, it comes in over 100 lbs once you include plug-ins.  It is the least capable of dual beam scopes in the 500 series, having only a single time-base connected to a single set of horizontal deflection plates, and a lone intensity control shared across both beams.
Nevertheless, I’m in the mood to work on a 500 series, and this was the only ‘normal’ one that’s currently the apartment; frankly I’m still too intimidated to work on the 519 or the 661.

Here it is on workbench B (aka the dining room table). On top of the power supply is my TU-75 variac which I was using to check the power supply on the 181 time mark generator (more on that later). Pardon the underwhelming photo quality, proper glamor shots are on the winter todo list.

IMG_5128.JPG

The vertical section is the same for both beams, and seems to be working well.  The timebase and triggering are are also in good shape, but the timebase is off by 3% – 8%  across almost all of the ranges.  Here’s output from my 181 timemark generator at 1ms.

IMG_5125-0.jpg

Parallax induced by the fact that the graticule is forward of the CRT surface makes it difficult to take an accurate picture, but the in the photo above, the left marker is aligned with the left graticule, and the right most marker is at the 2nd to last small tick; making this about 4% off.  Out of the factory, accuracy was within 1%, and I’d like to see about getting that restored.

The manual doesn’t have a section for calibration, it’s a separate document – both available on the Tek wiki, here. To calibrate the sweep, there’s a trim-pot on the back of the ‘Horizontal Display’ control, visible in the photo below, just behind the red test-lead.
IMG_5123.JPG

Update: What’s written below is mostly incorrect (though I’m leaving it for historical purposes).  I was able to get the timebase into calibration by following the instructions. Before calibrating the sweep, you’re supposed to calibrate the sweep magnification.  This seems completely counter-intuitive, and I never use the mag, so I didn’t bother with this step.  This turned out to be a giant mistake.  Once I did that, I was able to get the sweep calibrated quite easily, with the exception of .1/.2/.5ms, which are running a little slow.  That points to C180A, perhaps I’ll try to replace it another time.  For now, I’m calling this instrument done.


 

This pot is already as far as it will go, so no luck there.  Turning it the other way increases the error.  What’s interesting is this adjustment doesn’t actually alter the rate of the sweep ramp per se, it changes the amplitude.  There are trimmer caps for some of the faster sweep rates, but there’s nothing in the ramp generator to alter global timing.

The sweep signal is generated in the miller run-up circuit, given a variable DC offset to adjust the horizontal placement of the trace on the display, and then is turned into a differential signal and amplified before being sent to the horizontal deflection plates.
So what’s wrong with mine?  10ms worth of 1ms markers are being displayed in less than 10ms of divisions (1ms per division).  That means by the time the 10th marker goes off, the beam hasn’t yet gotten to the right spot on the screen.
What do we need to do to correct this? The slope of the ramp is determined by both the RC constant of the ramp generator, and the subsequent amplification stage, so it’s possible (within reason) to adjust for a slow ramp in the amplification stage. In the photo above, that’s what I’ve done:  By putting a 500k resistor in parallel with the 100k resistor that’s in series with the trim-pot, I was able to get the sweep into cal.  Is this cheating?  It feels like it.  I haven’t soldered it in yet, I’m just using a substitution box.  At higher sweep speeds, the capacitance of the leads have a significant impact on the waveform.

Screen Shot 2015-10-23 at 1.14.44 PM

It’s easy to make erroneous conclusions if you don’t understand the effect the controls have on the measurements, or don’t heed & grok all of the notes on the schematics.  While it’s possible to get the voltages listed on the schematics, if you have to deviate too far from the stipulated settings, or adjust other controls to their extremes, there’s probably something else wrong.  For example, I could get a few measurements within spec if I altered the sweep length beyond normal or listed ranges.  In another case, I measured the length of the sweep ramp at the cathode of V173, and found it to always be about 5-10% too long.  At first I thought this was indicative of the ramp running too slow and I’d found the issue, but then I realized that the sweep length control effects this, and it was only a coincidence.

I’ve gone back and forth between suspecting that the issue is in the ramp generator or the amplifier.  Like I said, the important thing is the slope of the ramp, and that can be adjusted (within reason) in either section.  All of the timing resistors & capacitors are within 1%

Here’s a bunch of the measurements.  It’s extremely challenging to find a 5% error among readings that are 10% off.
Screen Shot 2015-10-24 at 11.23.10 AM

This is getting annoying.  I may just solder that damn resistor in place and move on with my life.  More to come.

 

 

The Tektronix 519

So I got a 519.

IMG_3526

Here it is in a rather compromising position,  getting its caps checked.

“Turn your head and cough”IMG_3494.JPG

What’s a Tektronix 519?  Arguably the biggest, baddest, ballsiest oscilloscope ever fucking made.  Designed over 50 years ago, and built in extremely limited numbers through the 60s and early 70s, it was unsurpassed in its ability to capture extremely fast, transient or one-shot events, as well as repetitive signals of up to 1GHz,  in real time.

It is a machine unlike anything Tektronix had ever made, and due to the fact that it was the pet project of a senior engineer working free from the boundaries of the institution’s design team, it bared a only casual resemblance to the rest of their line-up.

Because the variac I have easy access to is only good for 5 amps or so (and gets a bit toasty at that),  I decided to just go for it, and hit it with full mains voltage.  Side note: I really need to get that 20A variac out of storage and wired up.  Nothing.

Checked the fuse, and there was a blown 4 amp slo-blow. Based on the markings, a 4 amp fuse suggests that the unit was being used at 240v, otherwise it would have had a 7 amp fuse for 120v  operation.  I carefully rolled the unit over to remove the bottom panel and expose the transformer wiring.  To my surprise, the transformer was wired for 120v operation.  I’m just preying no one tried to power this thing up on 240v, but I suspect not.  With the exception of a fine layer of dust, the interior is immaculate, and shows no signs of distressed components.  I’m also pretty sure I heard a brief ‘chunk’ of the blower try to start over the heavy click of the switch, so it is possible that it was I who blew the incorrectly installed fuse.

I’m taking this as a sign that I should do a more thorough inspection before firing this thing up for the first time in god only knows how many years.  With the chassis still rolled over (I would like to move this heavy, fragile, priceless relic as little as possible!), I went to work testing the supply caps who’s terminals are accessed on the underside of the chassis.  These are the ‘low voltage’ supplies, covering about 10 different ranges between -250v through +650v.  The high voltage supplies are -4,100v and +20,000v (!!).

Here’s my updated test & reforming rig.  My heathkit IT-28 capacitor checker, my trusty old Fluke 83 for measuring voltage across the cap, and a Bell & Howell multimeter for measuring current.  This is a kludge to say the least, but it’s what I had on hand.  A programmable, high voltage power supply tailored for cap reforming sounds like an interesting next project.  In the mean time,  this will have to do.

tek519testing

Here’s a closeup showing the charge line from the calibration step generator.  It’s in the way and making it difficult to disconnect the caps, lets see if we can remove it. IMG_3530.JPG

Removing the screws in the GR-874 connector…

IMG_3532.JPG

And loosening the clamps on the charge line…

IMG_3531.JPG IMG_3533.JPG

…let me carefully pull the charge line out through the front panel.

Here it is on my bench with its outer sleeve removed.IMG_3534.JPG

Here’s the coil that wraps around the charge line and triggers the reed relay within it.  It’s also worth noting that the bit of corrosion seen in the photo below represents the only bit of it I’ve found anywhere in here so far.IMG_3536.JPG

I started with C613, then moved on to C661,2,3, which is three 125uf caps in parallel.  I left the cap fully in-circuit for some of my early tests.

Once I disconnect the + side of the cap,  I connect the reforming circuit and start out at 50v and slowly work up to the rated voltage of the cap.  As the voltage increases,  it takes longer for the current to drop to an acceptable level.  As far as what is considered acceptable leakage current,  I’ve read a variety of opinions.  This site has a handy chart, which suggests that anything under a few mA is acceptable for these beastly 125uf, 450v caps.  I’ll move on to the next higher voltage when the leakage current drops below a mA,  then hold at the rated voltage until the current levels off.  So far, that’s been on the order of a few hundred uA, which is great for 45+ year old caps.

Ready for smoke testIMG_3545.JPG

so is Fire ChickenIMG_3546.JPG

YEAHHHH!IMG_3549.JPG

I was dorky enough to record a video.  Here it is in all of it’s dark, shaky, glory.  Pardon the occasional explicative, I was very excited.

I was able to get a free running trace, at all but one sweep speed.  The trace was crisp, and I was able to move it as expected via the positioning controls.  It is a shade of blue that I have never seen on a scope CRT before (P11 phosphor).  The 24kV accelerating potential made for a staticy, zappy sound when the relay kicked in to power the ‘low’ and high voltage supplies.

I hooked up the rate generator to the input.  While increasing the multiplier control, I heard a snap, and saw a white glow in V717, a 77http://paulcarbone.com/blog/wp-admin/post.php?post=409&action=edit34/6GE8 which is part of the +450V supply.  This I believe is completely unrelated, but obviously worth investigating.

Next Steps:

  1. Check V717 in my tube tester
  2. Check the output of the Rate Generator & Step generator with either my 7834, or my 661 (heheheee).
  3. See if I can get it to trigger on anything.

V717 is a 7734/6GE8, a combo triode + pentode.  What claims to be the latest settings sheet for my Heathkit TT-1 tube tester does not have this tube.  This is the second time that’s happened this month, and it’s starting to chap my ass.  I have a few other testers that are in need of resurrection, but I thought that the TT-1 was the most comprehensive.  I’ll have to double check next time I’m back up at ‘The House for Wayward Oscilloscopes’.  grrr….  Lets just put it back in and check the 450v line.  Perhaps this transient event I witnessed was not actually damaging to the tube.

Here’s the output of the Rate Generator on my 7603IMG_3552.JPG

With Function set to ‘Sync’, I was able to get this trace at 50nS / cm.  Very sporadic and required constant fiddling with the sync & vernier controls.
IMG_3553.JPG

After some more fiddling, I was able to get this trace at 10nS.  20nS is the one that doesn’t work.
IMG_3558.JPG
That’s insane.  I’ve never seen a trace that sharp before.  This thing is a monster.

7D01 Teardown

-IN PROGRESS POST-

DSC_0774

As I’ve previously mentioned, I’m in possession of no less then three, complete, working 7D01+ Display Formatter pairs (2 DF01s & 1DF02).  I got my original one working, keeping one as a maintenance spare / parts unit, and tearing down & stripping the final one.

It did take a moment of inward reflection before I decided to sacrifice a Tektronix product, but I decided:

  • It’s not particularly rare;  You’ll always see one or two on eBay.  Though they’re asking price is usually upwards of $200 (or sometimes more, which is just silly), they really only sell for about $50 – $100.  They were made from 1977 to 1985, though I don’t know how many were actually made.
  • It’s been an interesting circuit to get to know, and further poking will be educational.
  • It’s a trove of useful parts, including a beautiful ceramic & gold MC6800 CPU in the DF01, 20+ really nice toggle switches, tons of discrete logic, a dual concentric encoder, and some dual concentric pots.  I’ve already made a rudimentary 1Mhz clock circuit using some of it’s parts.
  • It’ll be fun, and I’ll take some pictures for here and the wiki.
  • Did I mention I have three of them?

The front panel circuit board is called out as the recognizer & trigger selector.  It also has the differential line receivers (Motorola MC10216) which are the input buffers for the P6451 probes.  The probes connect via 2 25-pin micro sub-D connectors.

 

 

7834 Readout repairs, continued

 

After my previous antics with the 7834 readout system, the outcome was:

  • U3401 Zeros logic & memory – tek part 155-0018 – the original culprit.  Replaced with one from ebay.
  • U3418 Column decoder –  tek part 155-0014, which I think I killed during repairs. Replaced with one from ebay.
  • U3477 7402 quad 2 input NOR gate – smoked during repairs.  Replaced with a 74ALS02 from stock.
  • It turned out U3232, the row data switch that cycles through the 8 plugin display channels was bad as well,  botching the readout in horizontal A bottom, and horizontal B top & bottom.   Replacement sourced from ebay.

So far, I’ve spent $50 on the scope, and another $50 or so on replacement chips.  Still not bad at all.

Here’s a few shots using the 7B85’s delta time feature, which lets you do fun things like measure frequency, rise-time, and pulse-width directly using the on-screen delay time read-out.  Thanks to Peter on the TekWiki who’s entry on the 7B85 pointed out these features.

These photos are of a rudimentary clock circuit I made using parts from the DF01

FullSizeRender FullSizeRender-1